Evaluation of corneal flap dimensions and blade quality using the automated
microkeratome SKBM
Viestenz A., Langenbucher A., Hofmann-Rummelt C., Viestenz A., Seitz
B.
Department of Ophthalmology, University Erlangen-Nürnberg, Schwabachanlage
6, 91054 Erlangen
Purpose: To evaluate flap dimensions and blade quality with repeated
blade use of an automated microkeratome.
Methods: The automated Summit Krumeich-Barraquer Microkeratome®
(SKBM, 160 µm plate, intended flap diameter 9 mm, cutting speed
up to 20000 osc/min), was utilized to perform a corneal hinged flap in
35 pig cadaver eyes. Seven blades were reused five times each. Flap diameter
was measured by planimetry and thickness was assessed by ultrasonic pachymetry.
Additionally, scanning electron microscopy (SEM) analysis of single and
multiple (up to five times) used blades was performed.
Results: With single use of the blade, mean central flap thickness
was 145±25 µm. The vertical / horizontal flap diameter was
9.0±0.03 mm / 8.6±0.03 mm. No correlation was found between
central thickness and horizontal (r=-0.15, p=0.4) or vertical flap diameter
(r=-0.04, p=0.83). No thickness gradient of the flap was observed in direction
from the incision (138±31 µm) to the flap hinge (130±30
µm). If the blade was used more than two times, the flap thickness
was smaller at the incision (157±34 µm vs. 124±20
µm; p=0.003) and the hinge (143±24 µm vs. 122±31
µm; p=0.04), but the central flap thickness remained unchanged.
The hinge size ranged from 2.5 mm to 6.0 mm (mean: 5.0±0.8 mm).
SEM analysis showed a progression of blade irregularities (up to 9.3 mm)
and larger tissue remnants on the multiple used blades.
Conclusion: Reproducible flap size of 9 mm diameter and thickness
of 145 µm in pig eyes can be obtained with single use of stainless
steel blades in the Summit Krumeich-Barraquer Microkeratome®. Because
of the peripheral thinning tendency of the flap and the increase of tissue
remnants repeated blade use can not be recommended.
|