Programm                 "Degeneration und Regeneration– Grundlagen, Diagnostik und Therapie"


Hotelbuchung
   Hotel Registration
Grußwort
   Welcome address
Beteiligte Gesellschaften
   Societies involved
DOG Information
   DOG Information
Eröffnung des Kongresses
   Opening Ceremony
Preise
   Awards
Ablauf der Tagung 2003
   General overview of congress
Lageplan der Räumlichkeiten
   Map of Congress Center
Wissenschaftliche Themen
   Scientific topics
Symposien
   Symposia
Wissenschaftliches Programm
   Scientific program
Posterpräsentationen
   Poster Presentation
Kurse
   Courses
Begleitende Veranstaltungen
   Accompanying program
Arbeitssitzungen
   Working sessions
Rahmenprogramm
   Social program
Allgemeine Informationen
   General Information
Autorenindex
   Index of Authors
Industrieaussteller
   Commercial exhibitors
Sponsoren
   Sponsors
Impressum



DOG Homepage


Abstract
Abstract

Comparison of Retinal Nerve Fiber Layer (RNFL) Measurements with Optical Coherence Tomography (OCT) and Retinal Thickness Analyser (RTA)

Thiel M. J., Neubauer A. S., Rombold F., Kampik A. 
University Eye Hospital, Ludwig-Maximilians Universität München

Purpose: Measurement of the RNFL can be of importance in detecting progressive glaucomatous damage. Several machines are well suited for measurements of the RNFL (Nerve Fiber Analyser (GdX), Heidelberg Retina Tomograph (HRT), OCT and RTA) and are commercially available. In this study we compared topography images from the RTA and the OCT as topography images can show defects at a glance and are superior for the clinically trained examiner.
Method: 20 eyes of 20 patients with ocular hypertension or glaucoma were included in this preliminary analysis. We performed RTA (RTA II, glaucoma analysis and retinal thickness map) and OCT (OCT II, concentric rings and RNFL map) examinations and compared the false colour coded images of the peripapillary RNFL.
Results: 4 patients showed a good correlation between the two methods, 10 patients showed moderate correlation and 6 patients showed no correlation
Conclusions: One reason for the unsatisfactory correlation of the two methods was a suboptimal centering algorithm by the RTA. Another reason was that the OCT shows sectorial analysis while the RTA results must be calculated from selected values across the posterior pole. A revision of the inbuilt data analysis of the two machines could yield a more comparable form of RNFL images.